DELRecon: Depth Electrode Reconstruction Toolbox for Stereo-EEG

Shameer Aslam,Qi Chen,Jiaxiang Zhang
DOI: https://doi.org/10.1007/978-3-031-67285-9_10
2024-01-01
Abstract:Many focal epilepsy patients resistant to medication are known to respond well to resective surgery. During the presurgical evaluation, stereo-EEG (SEEG) is often used to identify the epileptogenic zone. SEEG monitoring is an invasive procedure where patients are surgically implanted with depth electrodes into the brain. Existing methods for SEEG localization often lead to errors when multiple high-density electrodes are implanted close to each other. This study introduced an opensource MATLAB toolbox, DELRecon, for depth electrode reconstruction from post-implantation CT images. DELRecon included three processing modules: image processing, electrode localization and individual contact labelling. The electrode localization module used iterative clustering to trace each electrode and singular value decomposition to determine the electrode trajectories. The contact labelling module detected and labelled each contact through intensity-based clustering. The toolbox further included user interfaces for manual intervention when necessary. We evaluated DELRecon in a SEEG dataset, including 92 electrodes from 8 patients, totalling 1196 contacts. The toolbox accurately localized over 98% of all contacts. Our results suggested that DELRecon is an efficient solution for automatically localising SEEG electrodes and contacts.
What problem does this paper attempt to address?