Conformational Change-Based Fluorometric Aptasensor for Sensitive Cadmium (II) Detection in Fruits and Vegetables

Jiangxiong Zhu,Hao Yin,Yang Wang,Lumei Wang,Xueqing Geng,Yun Deng
DOI: https://doi.org/10.1039/d4ay01333f
IF: 3.1
2024-01-01
Analytical Methods
Abstract:Cadmium (Cd2+) is a highly toxic heavy metal that can accumulate in the human body through contaminated food and water, posing great health risks. In this study, a label-free fluorescent aptasensor based on SYBR Green I (SGI) for the rapid and sensitive detection of Cd2+ in food samples was designed. The aptasensor utilizes a Cd2+-specific aptamer (Cd-(21)) and its complementary strand (CSCd-(21)) to form a double-stranded DNA (dsDNA) structure in the absence of Cd2+. SGI intercalates into the dsDNA, resulting in a strong fluorescence signal. In the presence of Cd2+, the aptamer undergoes a conformational change, preventing the formation of dsDNA and leading to a decrease in fluorescence intensity. Under optimized conditions, the aptasensor exhibited a linear response to Cd2+ concentrations ranging from 0.11 to 157.37 ng mL(-1), with a limit of detection (LOD) of 0.07 ng mL(-1). The aptasensor demonstrated high specificity and was successfully applied to detect Cd2+ in fruits and vegetables, with satisfactory recovery rates (95-111%). The proposed aptasensor provides a promising tool for the rapid and sensitive detection of Cd2+ in food.
What problem does this paper attempt to address?