A Shock Advisory Algorithm Based on Hybrid Network for ECG Disturbed by Cardiopulmonary Resuscitation

Yining Chen,Yue Zheng,Yanqi Huang,Xiaomei Wu
DOI: https://doi.org/10.1016/j.bspc.2024.106502
IF: 5.1
2024-01-01
Biomedical Signal Processing and Control
Abstract:Background and Objective:During cardiopulmonary resuscitation (CPR), chest compressions generate significant artifacts, leading to pre-shock and post-shock pauses that are necessary for reliable rhythm analysis. To reduce the “hand-off” time and improve the accuracy of rhythm analysis during CPR, a shock advisory algorithm has been proposed.Methods:A dataset of 9696 segments of 8-s electrocardiogram (ECG) was collected from 260 out-of-hospital cardiac arrest patients and 67 patients from public datasets. We mixed ECG signals with CPR artifacts under four different signal-to-noise ratios to generate data segments contaminated with CPR artifacts. In total, 295 patient-cases were utilized for training and 32 for validation. We proposed a CPR artifact filtering algorithm based on variational mode decomposition and frequency peaks detection. The 8-s ECG segments before and after filtering were divided into sub-segments, and then fed into a hybrid network and fused by a correction module. The final classification results were obtained through a further voting mechanism.Results:The sensitivity, and specificity in this study were 94.13% and 95.24%, respectively. The specificity was highly competitive with existing algorithms.Conclusions:With further development, this algorithm could be integrated into automatic external defibrillators, thereby increasing the success rate of cardiac arrest interventions.
What problem does this paper attempt to address?