Construction and Evaluation of AuNPs Enhanced Electrochemical Immunosensors with [fe(cn)6]3−/4− and PPy Probe for Highly Sensitive Detection of Human Chorionic Gonadotropin

Qi Zhao,Zhencheng Chen,Chen-Wei Shan,Tao Zhan,Chen-Yang Han,Xiao-Zhen Feng,Heinz -Bernhard Kraatz
DOI: https://doi.org/10.1016/j.ijbiomac.2024.132963
IF: 8.2
2024-01-01
International Journal of Biological Macromolecules
Abstract:Human chorionic gonadotropin (HCG), a vital protein for pregnancy determination and a marker for trophoblastic diseases, finds application in monitoring early pregnancy and ectopic pregnancy. This study presents an innovative approach employing electrochemical immunosensors for enhanced HCG detection, utilizing Anti-HCG antibodies and gold nanoparticles (AuNPs) in the sensor platform. Two sensor configurations were optimized: BSA/Anti-HCG/c-AuNPs/MEL/e-AuNPs/SPCE with [Fe(CN)(6)](3-/4-) as a redox probe (1) and BSA/Anti-HCG/PPy/e-AuNPs/SPCE using polypyrrole (PPy) as a redox probe (2). The first sensor offers linear correlation in the 0.10-500.00 pg center dot mL(-1) HCG range, with a limit of detection (LOD) of 0.06 pg center dot mL(-1), sensitivity of 32.25 mu A center dot pg(-1)center dot mL center dot cm(-2), RSD <2.47 %, and a recovery rate of 101.03-104.81 %. The second sensor widens the HCG detection range (40.00 fg center dot mL(-1)-5.00 pg center dot mL(-1)) with a LOD of 16.53 fg center dot mL(-1), ensuring precision (RSD <1.04 %) and a recovery range of 94.61-106.07 % in serum samples. These electrochemical immunosensors have transformative potential in biomarker detection, offering enhanced sensitivity, selectivity, and stability for advanced healthcare diagnostics.
What problem does this paper attempt to address?