Establishment and characterization of noro-VLP measurement by digital ELISA

Takema Hasegawa,Yuriko Adachi,Kazumi Saikusa,Megumi Kato
DOI: https://doi.org/10.26434/chemrxiv-2024-5fqsq
2024-05-28
Abstract:Highly sensitive viral analytical techniques are essential tools for preventing the spread of infections. In this study, we established a digital enzyme-linked immunosorbent assay (ELISA) system to quantify norovirus proteins with high sensitivity. We used norovirus-like particles (noro-VLPs) as a surrogate for norovirus and constructed two digital ELISA systems using two different antibody pairs. The quantitative performance of the noro-VLP measurement using each digital ELISA system was evaluated. Both assay systems exhibited high sensitivity, good linearity, and high stability. The first system exhibited a limit of detection (LOD) of 87 pg/mL, correlation coefficient (R2) of 0.9984, inter-assay variation of 5.5 %, and intra-assay variation of 5.2 %. The second system exhibited an LOD of 19 pg/mL, R2 of 0.9984, inter-assay variation of 4.5 %, and intra-assay variation of 2.5 %. Comparison of the two systems using the same calibrant for unpurified and fractionated noro-VLPs revealed that the quantitative values for unpurified noro-VLPs were the same, whereas those for fractionated noro-VLPs were dramatically different. Our findings indicate that the reactivity to various components in the noro-VLP solution was altered depending on the different antibodies. Furthermore, our study highlights the importance of using appropriate calibrants, which contain the same ratio of components as the noro-VLP analyte, to afford accurate measurements.
Chemistry
What problem does this paper attempt to address?
The main objective of this paper is to establish a highly sensitive digital Enzyme-Linked Immunosorbent Assay (digital ELISA) system for quantifying norovirus proteins in norovirus-like particles (noro-VLPs) and to evaluate the performance of this system. Specifically, the researchers developed two different digital ELISA systems, both using different pairs of antibodies to detect norovirus-like particles. Through experimental validation, both systems demonstrated good linearity, high sensitivity, and stable performance. The first system had a limit of detection (LOD) of 87 pg/mL, a correlation coefficient (R²) of 0.9984, an inter-batch coefficient of variation of 5.5%, and an intra-batch coefficient of variation of 5.2%. The second system had an even lower detection limit of 19 pg/mL, also with a correlation coefficient of 0.9984, an inter-batch coefficient of variation of 4.5%, and an intra-batch coefficient of variation of 2.5%. Additionally, the study explored how different components of norovirus-like particle solutions affect quantitative results and emphasized the importance of ensuring that the calibrators and the samples to be tested have the same component ratios when using norovirus-like particles as calibrators for actual sample analysis to achieve accurate measurements. In summary, the problem this study aims to address is the development of a more sensitive and reliable protein quantification technique to more effectively monitor and prevent norovirus infections.