Impact of blood volume, air exposure duration, transport duration, and testing delay on plasma total carbon dioxide in simulated open collections using microtainers

Michael S. Reid,Isolde Seiden Long,Lawrence de Koning
DOI: https://doi.org/10.1016/j.clinbiochem.2024.110831
IF: 3.625
2024-10-03
Clinical Biochemistry
Abstract:Background Exposing blood specimens to air reduces plasma total carbon dioxide (TCO 2 ). We evaluated the degree of TCO 2 reduction in simulated open collection of neonatal blood in BD microtainers® (microtainers), microtainer transport duration and delayed testing of open plasma aliquots. Methods Venous blood was aliquoted into open microtainers in a 3x4 factorial design to simulate combined effects of blood volume (0.2–0.6 mL) and air exposure duration (0–5 min), with blood drawn in vacutainers as a control. Separate effects of in-hospital transport duration (0–120 min; whole blood), off-site transport duration (0–24 h; centrifuged whole blood), and duration plasma aliquots remained open (0–120 min) were evaluated by repeated testing. Findings were analyzed using repeated-measures ANOVA and Student's T-tests. Results In the factorial experiment, mean plasma TCO 2 in all microtainers was on average 3.5 mmol/L lower than in vacutainers. Smaller blood volume but not greater air exposure duration significantly (p < 0.05) reduced TCO 2 . Mean TCO 2 in filled (0.6 mL; 1–5 min air exposure) microtainers was on average 2.9 mmol/L lower than in vacutainers. Simulated off-site transport of microtainers containing centrifuged whole blood significantly reduced TCO 2 (4 h; mean −1.5 mmol/L), as did delayed testing of aliquoted plasma (15 min; mean −1.3 mmol/L). Conclusions Plasma TCO 2 decreased with reduced microtainer blood volume, extended transport duration of centrifuged blood and testing delay of aliquoted plasma. To minimize TCO 2 reduction, microtainers should be fully filled and tested rapidly. Laboratories should also consider whether an interpretive comment, correction factor or separate reference intervals are appropriate for these tests.
medical laboratory technology
What problem does this paper attempt to address?