Robotic Repositioning of Photoacoustic Tomography for Reproducible Long-Term Monitoring of Peripheral Artery in Vivo

Yongjian Zhao,Handi Deng,Xianghu Yu,Yisong Zhao,Ao Xu,Yuhan Chen,Cheng Ma,Li Liu
DOI: https://doi.org/10.1109/tmrb.2024.3371542
2024-01-01
IEEE Transactions on Medical Robotics and Bionics
Abstract:This paper presents a robotic arm assisted photoacoustic tomography imager that employs a direct visual servoing-enabled repositioning strategy designed for reproducible surveillance of peripheral arterial disease (PAD) progression. The system ensures precise repositioning of the cross-sectional location of the vessel of interest, allowing consistent measurements to track PAD progression and response to treatment. This strategy overcomes the difficulty of obtaining cross-sectional images of the same vessel at different monitoring intervals due to operator variability. In addition, the acquisition of high-resolution, high-contrast vessel images helps to clearly delineate long-term changes in the vasculature. To evaluate the effectiveness of the proposed experimental configurations and algorithms, we performed two sets of vascular phantom experiments and one set of servo-imaging experiments on human lower extremity vessels. The experimental results show 100% similarity for all three sets of grid segmentation comparisons, while the pixel-wise comparison similarity is 99.3% for the first set, 97.4% for the second set, and 98.9% for the in vivo experimental set. These results are important for monitoring the progression of PAD and predicting the risk of cardiovascular disease.
What problem does this paper attempt to address?