Measurement of Microdisplacement by Spin Manipulated Magnetometer of Nitrogen Vacancy Centers with Magnetic Flux Concentration Structure

Xin Liu,Zhonghao Li,Hao Zhang,Hao Guo,Zongmin Ma,Huan Fei Wen,Xin Li,Jun Tang,Jun Liu
DOI: https://doi.org/10.1109/ted.2024.3359177
IF: 3.1
2024-01-01
IEEE Transactions on Electron Devices
Abstract:The diamond nitrogen-vacancy (NV) centers have stable spin-coherent properties at room temperature. They are easily initialized and read out, highly manipulable, and hold great promise for quantum precision measurements. In this article, we construct a magnetic flux concentration (MFC) structure combined with a quantum spin manipulated magnetometer of NV centers. This structure is further combined with a uniform spatial magnetic field gradient to achieve high-resolution microdisplacement (MD). MD measurement with and without MFC is realized using Ramsey’s principle. The sensitivity of MD measurement with MFC is 150.84 mV/mm, and the experimental test range is 105–107 mm. The resolution of MD detection has been experimentally verified to be 1.95 $\mu $ m without MFC and 49 nm with MFC. The resolution of MD detection has been improved by a factor of 39.8. The displacement shot-noise-limited sensitivity of the magnetometer can reach 0.21 nm/Hz $^{\text{1/2}}$ . The results demonstrate that utilizing quantum spin manipulation and the MFC effect enables achieving both high resolution and rapid detection of MD for NV centers in systematic diamonds. This offers a novel approach for high-resolution MD measurement.
What problem does this paper attempt to address?