A Paired Phase and Magnitude Reconstruction for Advanced Diffusion-Weighted Imaging

Chen Qian,Zi Wang,Xinlin Zhang,Boxuan Shi,Boyu Jiang,Ran Tao,Jing Li,Yuwei Ge,Taishan Kang,Jianzhong Lin,Di Guo,Xiaobo Qu
DOI: https://doi.org/10.1109/tbme.2023.3288031
IF: 4.6
2023-01-01
IEEE Transactions on Biomedical Engineering
Abstract:Objective: Multi-shot interleaved echo planer imaging (Ms-iEPI) can obtain diffusion-weighted images (DWI) with high spatial resolution and low distortion, but suffers from ghost artifacts introduced by phase variations between shots. In this work, we aim at solving the ms-iEPI DWI reconstructions under inter-shot motions and ultra-high b-values. Methods: An iteratively joint estimation model with paired phase and magnitude priors is proposed to regularize the reconstruction (PAIR). The former prior is low-rankness in the k-space domain. The latter explores similar edges among multi-b-value and multi-direction DWI with weighted total variation in the image domain. The weighted total variation transfers edge information from the high SNR images (b-value = 0) to DWI reconstructions, achieving simultaneously noise suppression and image edges preservation. Results: Results on simulated and in vivo data show that PAIR can remove inter-shot motion artifacts very well (8 shots) and suppress the noise under the ultra-high b-value (4000 s/mm(2)) significantly. Conclusion: The joint estimation model PAIR with complementary priors has a good performance on challenging reconstructions under inter-shot motions and a low signal-to-noise ratio. Significance: PAIR has potential in advanced clinical DWI applications and microstructure research.
What problem does this paper attempt to address?