A New Method for Correction of Temperature Changes in Scintillation Detector Without Using a Radioactive Source
Tai Gao,Jiajun Wu,Lingli Li,Yulong Zhou,Haoyan Yang,ZhiZhen Ye,Danqi Wang,Zekun Zheng,Qin Jiang,Zizhen Xu,Lei He,Tao Zhou,Shiqi Feng,Yingrui Leng,Zhonghai Wang,Rong Zhou,Chaowen Yang
DOI: https://doi.org/10.1016/j.nima.2022.166908
2022-01-01
Abstract:Scintillation detectors are widely used in gamma-ray spectrum measurement because of their beneficial characteristics. However, scintillators and photoelectric conversion devices are sensitive to temperature changes, which leads to a drift in the energy spectrum. This paper presents a new method for calibrating a scintillation detector that uses a pulse from an ultraviolet light-emitting diode (UV-LED) to excite a BGO scintillator to generate fluorescence, and to control the drive voltage of the UV-LED to obtain different fluorescence peaks for energy spectrum calibration. Temperature experiments were carried out to obtain the curves of the energy spectrum drift and the 137Cs source for temperature changes in the range from -10 degrees C to +30 degrees C. With the UV-LED peak position at 10 degrees C selected as the standard peak, the relative deviation in the energy spectrum drift due to temperature changes was up to 24.41% without calibration, but this was reduced to 6.18% after using the UV-LED reference peak to calibrate. To optimize the calibration method, the pulse energy was calculated by integrating the falling edge of the waveform rather than the whole waveform, and this reduced the maximum relative deviation to 1.45%. Overall, this UV-LED calibration method effectively modifies the scintillation detector's energy spectrum drift and provides a new reference scheme for the online calibration of scintillation detectors.