Rapid Detection of CoxA24v from Acute Hemorrhagic Conjunctivitis by Real-Time RT-PCR

Jie Li,Fang Huang,Jing Guo,Cui Shu-juan,Jia Lei,Xi-tai Li,Xinyu Li,Quanyi Wang
2011-01-01
Abstract:Objective To further investigate the specificity,sensitivity,and accuracy of a established method of real-time RT-PCR to detect coxsackievirus A24 variant(CoxA24v),a cause of acute hemorrhagic conjunctivitis(AHC).Methods The specificity of the method was evaluated by detecting CoxA24v and several other enteroviruses.The CoxA24v strain was diluted in accordance with its TCID50 and the sensitivity of real-time RT-PCR was evaluated.The experiment was repeated and accuracy of the assay was evaluated.Clinical specimens from suspected cases were subjected to this assay.Results Real-time RT-PCR detected CoxA24v with a high level of specificity,and there were no cross reactions with common enteroviruses such as CoxA16 and EV71.The sensitivity of this test was up to 0.1 TCID50.Six different concentrations of CoxA24v were detected five times using this method.The results were very accurate.Clinical specimens from suspected cases were analyzed using this method,and positive PCR products were all sequenced.Results indicated CoxA24v and produced no false positives.Conclusion A real-time PCR assay was a quick,sensitive,specific,and accurate method that is suited to early diagnosis and differential diagnosis of AHC.