Study of Diamond Detector Application at the Front End of a High Intensity Hadron Accelerator

Guangyi Ren,Duohui He,Weimin Li,Yuxiong Li,Ming Zeng
DOI: https://doi.org/10.18429/jacow-ipac2015-mopty016
2015-01-01
Abstract:Diamond detectors function as beam loss or luminosity monitors for high energy accelerators, such as LHC, Babar, etc. Because of regular detectors insufficient protection of the front end, diamond detectors owning significant characteristics, like time resolution in the nanosecond range, radiation hardness and negligible temperature dependence. Thus, diamond detectors have been becoming promising candidates for detecting BLMs of fully super-conducting hadron accelerator, such as CADS, FRIB. In this paper, the sensitivity of diamond detectors was simulated by Monte Carlo program FLUKA and GEAN4. Meanwhile, we tested the performance of a new prototype of CVD diamond detector, and compared it with Si-PIN and Bergoz detectors at the storage ring of the HLS II. The results of the diamond detector were consistent with other two detectors well. More evaluation of diamond detectors in low energy radiation field are ongoing.