Finite Element Analysis of Contact Pressure Distribution for Domestic Artificial Knee Joint under Different Movements

Weiyi Chen
2013-01-01
Abstract:To study the effect of different movements on the contact pressure of various components of artificial knee joint in aligned mechanical axis,the processes of different movements are photographed including squatting,climbing stairs and walking,and then the kinematic data are analyzed in every motion state.Based on the drawings provided by domestic manufacturers,a 3D finite element model of artificial knee joint is built and calculated by applying movements and load on it.The results show that,for the contact pressure peak value,all the components are the minimum while double legs are standing,the femoral component is the largest while squatting 90°,the tibial cushion component is the largest while squatting 120°,and all the components are between the peak values in squatting 90° and 120°while climbing stairs;in walking process,the peak values for every component decrease successively in the motions of single leg standing,toe off and heel strike.It shows that the domestic artificial knee joint is in accord with the anatomical structure of Chinese knee joint,the peak values of contact pressure on medial and lateral sides are moderate and the contact area is larger,and the artificial knee joint can meet the clinical requirements.
What problem does this paper attempt to address?