The Value of Dual Energy Spectral CT in Reducing Metal Artifacts Caused by Metallic Implants after Scoliolosis Operation in Children

Jian Sun
2013-01-01
Abstract:Objective To investigate the value of dual energy spectral CT in reducing metal artifacts caused by metallic implants after scoliolosis operation in children.Methods Eight patients after scoliolosis operation were examined with dual energy spectral CT imaging mode.Three types of images were reconstructed: polychromatic 140 kVp(QC),monochromatic(MONO) images from 80-140 keV and MONO with metal artifact reduction system(MONO+MARs).For eight patient,ten positions with metallic implants were selected,and for every position,two image slices with the highest metal artifacts and one image slice with the least metal artifacts were selected.For every image slice,two regions of-interest(ROI)(on soft tissue inside and outside of vertebral canal) were selected.One experienced radiologists independently assessed image quality,measured standard deviations(SD) in ROI,and calculated artifact index(AI).Image quality scores were analyzed by Wilcoxon Signed Rank test,and artifact indexes were analyzed by Student t test.Results Image quality improved as the energy level(keV) increased on the MONO and MONO+MARs images.At about 120-140 keV,lowest SD and AI were obtained on MONO and MONO+MARs images for areas outside and inside the vertebral canal,respectively.There was significant difference of SD and AI outside the vertebral canal between MONO images and QC images(P0.05).There was no significant difference of these values between MONO+MARs images and QC images(P0.05).For the soft tissue inside vertebral canal,there were significant differences of SD and AI between MONO and QC images,and between MONO+MARS and QC images(all P0.05).There was no significant difference between MONO and MONO+MARs images(P0.05).Conclusion The image quality in the area with metal implants after scoliolosis operation was improved by monochromatic images with spectral CT imaging both inside and outside vertebral canal.For areas inside vertebral canal,the use of monochromatic images with metal artifact reduction system further improved image quality.
What problem does this paper attempt to address?