Applications of a High-Capacity Tensiometer for Direct Measurement of Suction

Rui Chen,Moe Key
IF: 2.4
2013-01-01
Journal of Hydraulic Engineering
Abstract:Only matric suctions below 100kPa can be measured by conventional tensiometers. When the measured suctions are larger than 100kPa,the conventional tensiometers fail due to the cavitation in measurement system. A new high-capacity tensiometer,which is able to directly measure matric suctions greater than 100kPa,was developed in this study on the basis of a general pressure transducer made in China. Suction measurement on soil specimens shows that the suction values measured by the new tensiometer are consistent with those pre-applied on the soil specimens by using the axis-translation technique. This comparison validated the response speed and accuracy of the tensiometer within the suction range of 400kPa. The new tensiometer was also used to measure the soil-water characteristic curve along a drying-wetting cycle within the suction range from 0 to 500kPa. By comparing with the results from the pressure plate test,the accuracy and endurance of the tensiometer and its applicability to drying-wetting cycle were further verified. Finally,the new tensiometer was used to monitor suction in the field and the monitoring lasted for 25 days,thereby justifying long duration and quick response of the tensiometer for suction measurement. The above applications suggest that the newly developed high-capacity tensiometer performs well regarding response,reliability,endurance and stability.
What problem does this paper attempt to address?