Reagentless Amperometric Immunosensor Based on Human Immunodefficiency Virus Diagnosis Marker Glycoprotein 160 Antibody Coated Gold-magnetic Particles Modified Electrode

Lu Wang
2009-01-01
Abstract:A novel amperometric immunosensor for human immunodefficiency virus(HIV)glycoprotein 160(gp160)based on nanoAu/Fe3O4/Chitosan(chit)composite film as immobilization matrix was investigated.Firstly,the complex [Cu(bpy)2(H2O)]2(C6H8O4)2(ClO4)4(CuL)was prepared and adsorbed on the surface of glassy carbon electrode(GCE),then nanoAu/Fe3O4/Chitosan composite film was modified on the surface of GCE|CuL and HIV gp160 antibody was furthermore immobilized on the surface to form one gp160 immunosensor(GCE|CuL/nanoAu/Fe3O4/Chit/anti gp160).CuL can be employed as an electron transfer mediator for catalytic activity detection of H2O2.After the immunosensor was incubated with HIV gp160 solution at 37 ℃ for 30 min,the electron transfer access of CuL to H2O2 is partly inhibited,which led to a linear decrease of the catalytic efficiency to H2O2 by the oxidation of immobilized CuL by H2O2 at-300 mV in pH 7.0 PBS.Under the optimal conditions,the linear range of gp160 by the immunosensor is 1-400 μg/L and the detection limit is 0.5 μg/L at 3 times noise.The immunosensor shows more sensitivity and accuracy than that only modified by gp160 antibody coated nano Au/chitosan film.This method simplifies the detection system without adding electron transfer media to solution and shortens the analytical time with one-step immunoassay,thus would be valuable for clinical immunoassay for HIV gp160 in AIDs patients' serums
What problem does this paper attempt to address?