Determination of Perfluorinated Compounds in Ambient Air by Accelerated Solvent Extraction Combined with High Performance Liquid Chromatography-Tandem Mass Spectrometry

Yi Yang
2010-01-01
Abstract:A method for the determination of perfluorinate compounds ( PFCs) in ambient air by accelerated solvent extraction ( ASE) combined with high performance liquid chromatography-tandem mass spectrometry ( HPLC-MS/MS) was developed. The air samples were collected with Anderson high-volume sampler model PS-1. The gas phases were mainly collected by polyurethane foam ( PUF) plug and the particle phases were mainly collected by quartz fiber filter ( QFF) . Loaded PUF plugs and QFFs were extracted individually by ASE,each for 15 min 3 times with a mixture solvent of methanol/acetone ( 1∶1,V/V) at 100 ℃ and pressure of 15 MPa. The extracts were concentrated by a K-D concentrator,blown down with a gentle stream of nitrogen and exchanged into methanol. HPLC separation was performed by using an agilent zorbax SB-C18 column at gradient elution of 0. 05 mol/L ammonium acetate and acetonitrile. A triple quadrupole MS operating in multiple-reaction monitoring ( MRM) mode was used for PFCs quantitative analysis. The linearity ranges were from 0. 1 to 50 μg/L and the limited of determinations of the method ranged from 0. 034 to 0. 54 pg/m3. The method was applied to determine trace PFCs in the air samples collected in Guangzhou in May-June 2007,the concentrations of total PFCs ranged from 5. 2 to 37. 4 pg/m3 in gas phase and from 11. 3 to 74. 3 pg/m3 in particle phase. The method is sensitive,accurate and suitable to monitor trace PFCs in ambient air.