Hemorrhage Detection by Magnetic Induction Tomography with Helmholtz Coil

Zilong Zhu,Chao Tan,Yixuan Chen,Feng Dong
DOI: https://doi.org/10.23919/CCC58697.2023.10240606
2023-01-01
Abstract:Due to the nonradioactive, noninvasive, low cost and portable features, magnetic induction tomography (MIT) is a promising technology in medical imaging. The conventional MIT sensor array generates divergent excitation magnetic field, resulting in poor sensitivity intensity and uniformity. To improve the intensity and uniformity of the sensitivity in MIT for hemorrhage detection, a local MIT method with Helmholtz coil is proposed. The Helmholtz coil provides a uniform background magnetic field in the measured region. To obtain sufficient independent measurements for hemorrhage reconstruction, the open planar MIT sensor array with gradiometers is applied in combination with the Helmholtz coil. The structure and excitation-detection strategy of the MIT sensor array are designed. The distribution of the magnetic field, phase sensitivity and the phase response of hemorrhages under the designed MIT sensor array are analyzed. The results indicate that the open planar MIT sensor array with Helmholtz coil can effectively improve the intensity and uniformity of the phase sensitivity. In addition, it can also enhance the intensity of the hemorrhage signals for different depths and volumes of the hemorrhages. The proposed method provides a solution to improve the detection accuracy and the detection depth of the hemorrhage by MIT.
What problem does this paper attempt to address?