Low-Pressure Microwave Plasma Sterilization of Slender Tube Inner Surfaces

Miaomiao Xu,Huacheng Zhu,Yang
DOI: https://doi.org/10.1109/iwem58222.2023.10234907
2023-01-01
Abstract:To tackle the current issue of incomplete sterilization of slender medical devices, a microwave plasma sterilization device has been designed for studying the sterilization of the inner surface of slender tubes. The experimental device showed a good sterilization effect when the microwave frequency was 2.45 GHz, the microwave power was set to 200 W, the processing time was 10 minutes, and the oxygen-to-argon ratio was greater than 14.45%. The findings indicate a strong correlation between the sterilization effect of microwave plasma and the gas composition. The inclusion of oxygen emerges as a crucial factor in achieving microwave plasma sterilization. The experimental device successfully sterilizes the inner surface of the slender tube, thereby highlighting the extensive potential of microwave plasma.
What problem does this paper attempt to address?