Development and Validation of Data-Level Innovation Data-Balancing Machine Learning Models for Predicting Optimal Implantable Collamer Lens Size and Postoperative Vault.

Heng Zhao,Tao Tang,Yuchang Lu,Xuewei Li,Liyuan Sun,Sitong Chen,Lu Ma,Yan Luo,Kai Wang,Mingwei Zhao
DOI: https://doi.org/10.1007/s40123-023-00841-7
2024-01-01
Ophthalmology and Therapy
Abstract:There are only four sizes of implantable collamer lens (ICL) available for selection, which cannot completely fit all patients as a result of the discontinuity of ICL sizes. Sizing an optimal ICL and predicting postoperative vault are still unresolved problems. This study aimed to develop and validate innovative data-level data-balancing machine learning-based models for predicting ICL size and postoperative vault. The patients were randomly assigned to training and internal validation sets in a 4:1 ratio. Feature selection was performed using analysis of variance (ANOVA) and Kruskal–Wallis feature importance methods. Traditional linear regression model and machine learning-based models were used. The accuracy of models was assessed using the area under the curve (AUC) and confusion matrix. A total of 564 patients (1127 eyes) were eligible for this study, consisting of 808 eyes in the training set, 202 eyes in the internal validation set, and 117 eyes in the external validation set. Compared with the traditional linear regression method, the machine learning model bagging tree showed the best performance for ICL size selection, with an accuracy of 84.5