The Wavefront Analyzer for Measurement of Scotopic Pupil Size Before Refractive Surgery

Hu Chen
2009-01-01
Abstract:Objective To evaluate the reliance and repeatability of measurement of pupil size before refractive surgery using allegretto wavefront analyzer. Methods The pupil size of 100(200eyes) myopic subjects was measured under three illuminance levels using the Allegretto wavefront analyzer and compared with Haab ruler. Results The pupil size measured under three illuminance levels by Allegretto wavefront analyzer is respectively 7.42±0.92 mm(0.5 lux),6.87±0.82mm(1.5 lux)and 6.05±0.75 mm(6.5 lux).There are significant differences in the pupil size among three illuminance condition(P0.01).Under 0.5lux,the rang of pupil size measured by Allegretto wavefront analyzer is 5.12-9.87 mm.There are 40 eyes(20 %) more than 8 mm and 5 eyes(2.5%) more than 9 mm.Under 1.5 lux,there are no significant differences in the pupil size between the Allegretto wavefront analyzer and Haab ruler(P0.05). Conclusion The Allegretto wavefront analyzer can measure pupil size under different scotopic condition.It may be a better choice for refractive surgery.
What problem does this paper attempt to address?