Application of High Performance Liquid Chromatography(HPLC)in Detection of L-carnitine Levels of Plasma in Tilapia
Xiao TANG,Kai HUANG,Qiling YANG,Zhiyuan LUO,Linhua WU,Teng ZUO
DOI: https://doi.org/10.16378/j.cnki.1003-1111.2018.03.011
2018-01-01
Fisheries Science
Abstract:The effects of mobile phase,column temperature to chromatographic peak,method of sample handling and derivation reaction conditions on L-carnitine levels of plasma in GIFT tilapia Oreochromis niloticus with body weight of(82.71 ± 1.28)g,and the corresponding HPLC method was established un-der conditions of Kromasil Sil chromatographic column(250 mm×4.6 mm,5 microns),and mobile phase consisting of acetonitrile-citric acid buffer solution(90:10,V/V),0.25 mL per 1 L liquid phase joining triethylamine,flow rate of 1.2 mL/min,35 ℃ of column temperature,detection wavelength of 260 nm and 20 μL of sample.The 10% of bromobenzene acetyl bromide was used as derivative reagent at 70 ℃water bath for 90 minutes.Methodological investigation revealed that the corresponding HPLC method has high specificity,with the equation of linear regression and the correlation coefficient of y=0.08376 x+0.09663(n=6,r2=0.9995),the minimum detection limit of 1 μmol/L,the intra-day and inter-day varia-tion coefficients of 1.79% —5.56% and 9.83% —11.81%,respectively,the average recovery of 95.83%, and the 12 hours peak area relative standard deviation of 6.58%.The findings indicate that the established HPLC method can be used as an assay for detection of L-carnitine in fish.