SPE-HPLC/FLD Method for the Determination of Four Estrogens in Water Samples

Wanting Ling
2012-01-01
Abstract:A solid phase extraction(SPE)-high performance liquid chromatography(HPLC)-fluorescence detection(FLD)method was established for the simultaneous determination of four estrogens(estriol,17β-estradiol,17а-ethynylestradiol,and bisphenol A)in water.Water samples was concentrated using a C18 solid phase extraction column,which was then consequently eluted with 5.00 mL ultrapure water and 15.00 mL methanol.Eluent was dried by nitrogen blow.The residue was then dissolved with 50% methanol solution,and determined by HPLC-FLD with the mobile phase of methanol/acetonitrile/wate(rvolume ratio of 25:30:45)at 1.0 mL.min-1,Inertsil ODS-SP-C1(8150 mm× 4.6 mm,5μm)reversed phase column at 40 ℃,excitation and emission of 280 nm and 310 nm,and injection volume of 20 μL.The linear range for the detection of test four estrogens was 5.00~1 000.00 μg.L-1 with correlations(r)higher than 0.999 9.Four estrogens detection limits of HPLC-FLD were 0.107~0.271 μg.L-1,the detection limits of SPE-HPLC/FLD were 0.214~540 ng.L-1.The recoveries of the four estrogens with SPE-HPLC/FLD were more than 85%,except 17а-ethynylestradiol of 60%,with relative standard deviations(RSD)(n=5)less than 4%.The method has high sensitivity,low detection limit,good reproducibility and precision,and may be used for the determination of trace amounts of estrogen in water environment.
What problem does this paper attempt to address?