Model Error Correction for Linear Methods in PET Neuroreceptor Measurements

Hongwei Guo,Rosemary A. Renaut,Kewei Chen,Eric M. Reiman,Samaritan Positron
2008-01-01
Abstract:Graphical analysis methods are widely used in positron emission tomography quantification because of their simplicity and model independence. But they may, particularly for reversible kinetics, lead to bias in the estimated parameters. The source of the bias is commonly attributed to noise in the data. Assuming a two-tissue compartmental model for neuroreceptor studies, we investigate the bias that originates from model error. This bias is an intrinsic property of the simplified linear models used for limited scan durations, and it is exaggerated by random noise and numerical quadrature error. Conditions are derived under which Logan’s graphical method either over- or under-estimates the distribution volume in the noise-free case. The bias caused by model error is quantified analytically. The presented analysis shows that the bias of graphical methods is inversely proportional to the dissociation rate. Furthermore, visual examination of the linearity of the Logan plot is not sufficient for guaranteeing that equilibrium has been