High-Precision Thermoelectric Measurement Device in Water-Cooled Magnet under High Magnetic Fields

QIAN Yingcai,ZHANG Yong,LI Qi,PAN Senyang,ZHANG Jinglei,PI Li
DOI: https://doi.org/10.13380/j.ltpl.2022.06.005
2022-01-01
Abstract:Recently, topological materials have stimulated unprecedented research interest in condensed matter physics.In the presence of a perpendicular magnetic field and a longitudinal thermal gradient, the diffusion of carriers can produce a longitudinal electric field and a transverse electric field, which are defined as thermoelectric effects. In principle, thermoelectric effects are sensitive and promising probes to novel quantum states and exotic physical effects for those topological materials. However, the studies of thermoelectric properties in ultra-high magnetic fields are rare. In this work, aiming at the problems of fast field-sweeping rate and large mechanical vibration, we have designed and constructed a high-precision measurement device for detecting the thermoelectric effect in water-cooled magnet up to 32 T. Meanwhile, the reliability of our measurement system is further confirmed by measuring the thermoelectric effects for several topological materials.
What problem does this paper attempt to address?