The Optimum Resistance of Small Intestine Fusion by Pulse Source

Yifan Wang,Kefu Liu,Xiaowei Xiang,Caihui Zhu,Hao Wang
DOI: https://doi.org/10.21203/rs.2.21965/v1
2020-01-01
Abstract:Abstract Background The radiofrequency (RF)-induced intestine-fusion, as a replacement method for traditional suture in surgery, has been studied for years. The present study reports the optimum resistance of small intestine fusion. As the feedback signal, resistance will be the indicator of the fusion completion for device design of intestine-fusion, and in-depth study of microstructure change. Materials and methods A self-design pulse source is used for small intestine fusion with adjustable voltage, duty ratio, frequency and output time. In this study, the frequency is 440 kHz; voltage, output time and compression pressure (CP) on the small intestine are independent variables for the experiment. The differences of hematoxylin-eosin (H&E) staining slices are discussed. The real time current is measured and recorded during the fusion for calculation of resistance. The burst pressure (BP) is measured with a pressure gauge and a peristaltic pump after the fusion completion. Results The highest BP is 38.9 mmHg with CP of 900 kPa, voltage of 50 v and time of 5 s. The optimum resistance is 71.1 Ω. Conclusions The quality of fusion is a result of combined influence of CP, voltage and time. The optimum resistance of 71.1 Ω is proposed for the highest BPs and automatically fusion.