Potentiometric Immunosensor Based on Immobilization of Hepatitis B Surface Antibody on Platinum Electrode Modified Silver Colloids and Polyvinyl Butyral As Matrixes

Dianping Tang,Ruo Yuan,Yaqin Chai,Linyan Zhang,Jianyuan Dai,Yan Liu,Xia Zhong
DOI: https://doi.org/10.1002/elan.200403066
IF: 3.077
2005-01-01
Electroanalysis
Abstract:A highly sensitive immunosensor based on immobilization of hepatitis B surface antibody (HBsAb) on platinum electrode (Pt) modified silver colloids and polyvinyl butyral (PVB) as matrixes has been developed for potentiometric immunoanalysis to detect hepatitis B surface antigen (HBsAg) in this study. HBsAb molecules were immobilized successfully on nanometer-sized silver colloid particles associated with polyvinyl butyral on a platinum electrode surface. The modification procedure was electrochemically monitored by electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV). The HBsAb-silver-PVB-modified electrode exhibited direct electrochemical behavior toward HBsAg. The factors influencing the performance of the resulting immunosensor were studied in detail. More than 94.7% of the results of human serum samples obtained by this method were in agreement with those obtained by enzyme-linked immunosorbent assays (ELISAs). The resulting immunosensor exhibited a sigmoid curve with log HBsAg concentration, high sensitivity (39.8 mV/decade), wide linear range from 16.0 to 800 ng mL−1 with a detection limit of 3.6 ng mL−1, fast potentiometric response (<3 min) and long-term stability (>4 months). The response mechanism of the immunosensors was also studied with AC impedance techniques.
What problem does this paper attempt to address?