Ultrasensitive Sponge-Inspired Mxene@Ceo2 Detector for Glucose Detection in Sweat

Shuyue Jiang,Haifeng Zhang,Zhaoxin Li,Liang Chen,Liang Yin,Xiaowei Liu
DOI: https://doi.org/10.2139/ssrn.4328213
2023-01-01
Abstract:Painless glucose detection is the greatest wish of diabetic patients. To realize this wish, blood glucose detection can be replaced by sweat glucose detection. However, there is only a slight amount of glucose in sweat. In this study, an ultrasensitive biomimetic glucose detector is proposed, inspired by the predator process of sponge in sea. The MXene (Ti3C2Tx) @CeO2 detection electrode employed the “laminae-trestle-laminae” (L-T-L) structure of sponge to adsorb and transport glucose molecules in sweat. CeO2 grafted onto the MXene surface can detect the glucose by the redox reaction of CeO2 and glucose. The glucose detector with enhanced sensitivity (801.27 µA·µM-1·cm-2), low limit of detection (LOD=0.004 μM), linear detection range (LDR=0.01-2.5 μM) can satisfy glucose detection in human sweat. The best recovery of this biomimetic detector in artificial sweat is 99.6%. In addition, the detector not only completes the detection of sweat glucose in diabetic patients for five consecutive days, but also can detect the glucose content in the daily drinks of patients. Our findings demonstrate that high potential of sponge-inspired MXene@CeO2 glucose detector for biomedical applications on a daily basis.