Design and Evaluation of a Novel Coring Atherectomy System for Arterial Calcified Plaque

Zecui Zeng,Chen,Qiang Ye,Geng Huang,Dongchao Yang
DOI: https://doi.org/10.2139/ssrn.4297905
2022-01-01
SSRN Electronic Journal
Abstract:Arterial stenosis caused by calcified plaque that grows inside the vasculature would decrease the blood flow, many medical methods are used to remove the plaque to improve the blood flow. Atherectomy is an efficient way to remove the plaque with lower risk of relapse. However, a great deal of fragments are generated by the conventional atherectomy product during cutting the calcified plaque off, and these fragments should be collected by other surgery equipment. A novel coring atherectomy system (NCAS) is proposed to improve this situation, the NCAS consists of a coring rotablation, soft drive shaft with center hole and negative pressure unit. The maximum work torque of the soft drive shaft is calculated, and the effect of the rotablation rotation speed on removing the plaque is evaluated. The artery phantom is simulated by different size of silicon tube, and the plaque phantoms with different stenosis grade are fabricated by gelatin, CaCO3 and wax. Finally, the experiment setup is built up to evaluate the performance of NCAS. The experiment results show that the NCAS could remove the plaque efficiently, and more than 80% the fragment is stored in the catheter which would be expelled outside the body directly. Otherwise, we found that the soft plaque could be sucked into the catheter without atherectomy by a negative pressure unit.
What problem does this paper attempt to address?