Construction of a CuNPs-aptasensor for rapid detection of okadaic acid

YAN Xiao-Chen,LI Ling,QI Xiao-Yan,LI Jiao,WANG Sai,MAO Xiang-Zhao
DOI: https://doi.org/10.19812/j.cnki.jfsq11-5956/ts.2023.07.053
2023-01-01
Abstract:Objective To construct a CuNPs-aptamer sensor for the rapid detection of okadaic acid(OA).Methods Based on the previous research, the label-free fluorescent CuNPs-aptasensor was constructed according to the molecular binding mechanism between OA and its aptamer, along with the fluorescence signal changes triggered by the terminal conformational change of the aptamer. Highly sensitive detection of OA was achieved according to the suppression of the fluorescence signal changes caused by the OA and aptamer binding. Moreover, detailed evaluation was performed on the sensitivity, selectivity, accuracy, etc.. Results Linear detection range of the constructed CuNPs-aptasensor ranged from 75.0 to 2400.0 nmol/L, and the limit of detection was as low as 1.045 nmol/L.Moreover, high selectivity and accuracy was realized with the CuNPs-aptasensor analyzing shellfish samples.Conclusion The constructed aptasensor can achieve highly sensitive and specific detection of OA, and the study provides reference for the simple and rapid detection of other food targets.
What problem does this paper attempt to address?