Temperature Changes of Renal Calyx During High-Power Flexible Ureteroscopic Moses Holmium Laser Lithotripsy: a Case Analysis Study

Zhiyue Wu,Jingchao Wei,Chengfang Sun,Qi Huangfu,Bohan Wang,Guangyi Huang,Kai Qi,Zeping Jiang,Wuan Zhu,Jiaming Wen
DOI: https://doi.org/10.1007/s11255-023-03611-3
IF: 2.2662
2023-01-01
International Urology and Nephrology
Abstract:The risk of thermal damage increases with the introduction of high-power lasers during holmium laser lithotripsy. This study aimed to quantitatively evaluate the temperature change of renal calyx in the human body and the 3D printed model during high-power flexible ureteroscopic holmium laser lithotripsy and map out the temperature curve. The temperature was continuously measured by a medical temperature sensor secured to a flexible ureteroscope. Between December 2021 and December 2022, willing patients with kidney stones undergoing flexible ureteroscopic holmium laser lithotripsy were enrolled. High frequency and high-power settings (24 W, 80 Hz/0.3 J and 32 W, 80 Hz/0.4 J) were performed for each patient with room temperature (25 °C) irrigation. In the 3D printed model, we studied more holmium laser settings (24 W, 80 Hz/0.3 J, 32 W, 80 Hz/0.4 J and 40 W, 80 Hz/0.4 J) with warmed (37 °C) and room temperature (25 °C) irrigation. Twenty-two patients were enrolled in our study. With 30 ml/min or 60 ml/min irrigation, the local temperature of the renal calyx did not reach 43 °C in any patient under 25 °C irrigation after 60 s laser activation. There were similar temperature changes in the 3D printed model with the human body under the irrigation of 25 °C. Under the irrigation of 37 °C, the temperature rise slowed down, but the temperature in the renal calyces was close to or even exceeded the 43 °C at the setting of 32 W, 30 ml/min and 40 W, 30 ml/min after continuing laser activation. In the irrigation of 60 ml/min, the temperature in the renal calyces can still be maintained within a safe range after continuous activation of a holmium laser up to 40 W. However, continuous activation of 32 W or higher power holmium laser in the renal calyces for more than 60 s in the limited irrigation of 30 ml/min can cause excessive local temperature, in such situation room temperature perfusion at 25 ℃ may be a relatively safer option.
What problem does this paper attempt to address?