Design of a Lumbar Muscle Fatigue Monitoring System Based on Electrical Impedance Myography

Xu Li,Jingting Shi,Yueming Gao,Pan Xu
DOI: https://doi.org/10.1109/apcap56600.2022.10069995
2022-01-01
Abstract:In the process of lumbar muscle injury prevention as well as rehabilitation training, there is still a lack of means to detect and evaluate changes in lumbar muscle status in real time. To address this challenge, an Electrical Impedance Myography(EIM)-based lumbar muscle fatigue monitoring system is designed in this paper to track the changes of lumbar muscle status in real time. The system includes an impedance acquisition module with AD5933 as the core, an STM32 master control module, a Bluetooth communication module and an Android smartphone APP module. Muscle impedance measurements were performed on the lumbar erector spinae muscles of subjects by establishing an in vivo experimental platform for lumbar muscle impedance measurement, and the results were compared with those obtained by a commercial body composition analyzer. The results show that the system can continuously measure the electrical impedance of human lumbar muscles, the measured impedance values tend to decrease with the gradual fatigue of the lumbar muscles. The root-mean-square error (RMSE) of the results is less than 5% compared with that of commercial instruments, and the accuracy can meet the requirements of muscle fatigue assessment. In addition, the system designed in this paper can be operated easily and portably and monitored in real time by cell phone APP, which is a guide and reference for the design and development of new portable testing devices.
What problem does this paper attempt to address?