Endoscopic en-face optical coherence tomography and fluorescence imaging using correlation-based probe tracking
Manuel J. Marques,Michael R. Hughes,Adrián F. Uceda,Grigory Gelikonov,Adrian Bradu,Adrian Podoleanu
DOI: https://doi.org/10.1364/boe.444170
IF: 3.4
2022-01-14
Biomedical Optics Express
Abstract:Forward-viewing endoscopic optical coherence tomography (OCT) provides 3D imaging <i>in vivo</i>, and can be combined with widefield fluorescence imaging by use of a double-clad fiber. However, it is technically challenging to build a high-performance miniaturized 2D scanning system with a large field-of-view. In this paper we demonstrate how a 1D scanning probe, which produces cross-sectional OCT images (B-scans) and 1D fluorescence T-scans, can be transformed into a 2D scanning probe by manual scanning along the second axis. OCT volumes are assembled from the B-scans using speckle decorrelation measurements to estimate the out-of-plane motion along the manual scan direction. Motion within the plane of the B-scans is corrected using image registration by normalized cross correlation. <i>En-face</i> OCT slices and fluorescence images, corrected for probe motion in 3D, can be displayed in real-time during the scan. For a B-scan frame rate of 250 Hz, and an OCT lateral resolution of approximately <math><mrow><mn>20</mn></mrow> <mspace></mspace> <mrow><mi>μ</mi></mrow> <mrow><mi>m</mi></mrow> </math> , the approach can handle out-of-plane motion at speeds of up to 4 mm/s.
radiology, nuclear medicine & medical imaging,biochemical research methods,optics