A Fast Calibration Method Based on Constraints for Magnetic Sensor Array for Positioning Capsule Endoscope.

Mingxin Ren,Xiaoyang Wu,Jiaole Wang,Shuang Song
DOI: https://doi.org/10.1109/robio55434.2022.10011667
2022-01-01
Abstract:The magnetic localization method is suitable for gastrointestinal endoscopy, and the magnetic sensor array can track the position and direction of the capsule in the body. In order to achieve an accurate result, the magnetic sensor needs to be calibrated in advance. However, the existing calibration methods universally demand a large number of samples with known magnet poses, and then use optimization algorithm to obtain calibrated coefficients. To simplify the calibration process, a fast calibration method is proposed in this paper, which does not require to know the specific information of the calibration points. By introducing a novel calibration tool to restrict the orientation of the magnet and the position of the rotating fixed point, the analytical solution of the sensitivity coefficient of the magnetic sensor can be calculated quickly. In this paper, the calibration coefficients under different magnetic rotation ranges have been calculated and compared, and the results verified the effectiveness of the proposed calibration method.