Error Identification and Compensation Regarding the Kinematic Parameter of the MD-PEF for Tibial Deformity Correction.

Guotong Li,Xiaoqiang Tang,Jianfeng Li,Mingjie Dong
DOI: https://doi.org/10.1016/j.compbiomed.2023.106813
IF: 7.7
2023-01-01
Computers in Biology and Medicine
Abstract:The correction accuracy of an external fixator is crucial to the treatment outcome of deformity correction and patient safety. In this study, the mapping model is established between the pose error and kinematic parameter error of the motor-driven parallel external fixator (MD-PEF). Subsequently, the kinematic parameter identifi-cation and error compensation algorithm of the external fixator is established based on the least squares method. An experimental platform based on the developed MD-PEF and Vicon motion capture system is constructed for kinematic calibration experiments. Experimental results show that the correction accuracy of the MD-PEF after calibration is as follows: translation accuracy dE1 = 0.36 mm, axial translation accuracy dE2 = 0.25 mm, angulation accuracy dE3 = 0.27 degrees, and rotation accuracy dE4 = 0.2 degrees. The accuracy detection experiment verifies the kinematic calibration results, which further validates the feasibility and reliability of the error identification and compensation algorithm constructed by the least squares method. The calibration approach used in this work also provides an effective way to improve the accuracy of other medical robots.
What problem does this paper attempt to address?