Pulse Transition Characterization from Electrocardiography and Photoplethysmography for Non-Invasive Blood Pressure Estimation.

Hazem Mohammed,Hao Wu,Guoxing Wang
DOI: https://doi.org/10.1109/ISCAS48785.2022.9937252
2022-01-01
Abstract:Pulse Arrival Time (PAT) is a significant pulsetransition based feature used for non-invasive Blood Pressure (BP) monitoring. Electrocardiography (ECG) and Photoplethysmography (PPG) are usually used to measure PAT. These signals have multiple characteristic points that can be used as references for measurement. However, based on maximizing feature-BP correlation, most research attempts are focused on the R-peak of ECG and the characteristic points of the rising curve of the PPG. In this paper, to properly characterize the pulse transition, we study the T-wave of the ECG and the falling curve of the PPG to extract different reference points for PAT measurement. Such features hold significant information about the heart and the aorta, which are directly involved in blood flow and pressure. Consequently, these features can improve BP estimation significantly despite having a lower correlation with BP. The study is assessed based on 2152 subjects from VTLdatabase, an opensource surgical database of vital signals. Previous work based on this dataset yielded a root mean squared error (RMSE) of 11/6.25 mmHg for Systolic/Diastolic BP (SBP/DBP); based on traditional PAT features. By incorporating the newly proposed features, the estimation RMSE was lowered to 6.47/3.55 mmHg for SBP/DBP.