Optimal hippocampal targeting in responsive neurostimulation for mesial temporal lobe epilepsy

Henry M. Skelton,Katie Bullinger,Faical Isbaine,Jonathan C. Lau,Jon T. Willie,Robert E. Gross
DOI: https://doi.org/10.3171/2024.2.jns231732
IF: 5.408
2024-04-01
Journal of Neurosurgery
Abstract:OBJECTIVE The aim of this study was to identify features of responsive neurostimulation (RNS) lead configuration and contact placement associated with greater seizure reduction in mesial temporal lobe epilepsy (MTLE). METHODS A single-center series of patients with MTLE treated with RNS were retrospectively analyzed to assess the relationship between anatomical targeting and seizure reduction. Targeting was determined according to both the preoperatively conceived lead configuration and the actual placement of RNS contacts. Three lead configurations were used: 1) single bilateral, with 1 depth lead in each hippocampus; 2) single unilateral, with 1 hippocampal depth lead and another implant outside the mesial temporal lobe; and 3) dual unilateral, with 2 leads in 1 hippocampus. Contact placement on postoperative imaging was measured according to the number of hippocampal contacts per targeted hippocampus (contact density) and per patient (contact count), distribution throughout the hippocampus, and proximity to the anteromedial hippocampus. RESULTS Dual unilateral lead placement resulted in significantly higher hippocampal contact density compared with the single hippocampal approaches, but only showed a nonsignificant trend toward a higher rate of response. However, those patients with more than 4 contacts in a single hippocampus, achievable only with dual unilateral leads, had a significantly higher rate of response. The higher likelihood of response was poorly explained by more widespread hippocampal coverage, but well correlated with proximity to the anteromedial hippocampus. CONCLUSIONS Dual unilateral hippocampal implantation increased RNS contact density in patients with unilateral MTLE, which contributed to improved outcomes, not by stimulating more of the hippocampus, but instead by being more likely to stimulate a latent subtarget in the anterior hippocampus. It remains to be explored whether a single electrode targeted selectively to this region would also result in improved outcomes.
surgery,clinical neurology
What problem does this paper attempt to address?