An Ultra-Sensitive Core-Sheath Fiber Strain Sensor Based on Double Strain Layered Structure with Cracks and Modified MWCNTs/silicone Rubber for Wearable Medical Electronics

Shunge Wu,Ping Liu,Wei Tong,Junliang Li,Guangyuan Xu,Fei Teng,Jian Liu,Han Feng,Ruohai Hu,Austin Yang,Caixia Liu,Kun Xing,Xiaoping Yang,Helei Tian,Aiguo Song,Xiaoming Yang,Ying Huang
DOI: https://doi.org/10.1016/j.compscitech.2022.109816
IF: 9.1
2022-01-01
Composites Science and Technology
Abstract:Flexible resistive sensors have attracted widespread attention in the field of wearable electronics. Inspired by the single-layer crack structure which has ultra-high sensitivity, a double strain layered crack structure is proposed to expand the maximum working strain (30%) of flexible sensor while maintaining the ultra-high sensitivity (GF = 2165 at 0-30% and 7381 at 25-30%). MWCNTs modified by 3-Aminopropyltriethoxysilane were filled into silicon rubber to form nanocomposites with fast response times (42 ms), high repeatability (>1000) and low hysteresis (7.08%). Furthermore, the respiratory rate, heart rate and blood pressure can be measured by the proposed sensor. The average errors of the respiratory rate and heart rate are less than 2% and 3% respectively. The mean error of systolic blood pressure, diastolic blood pressure and mean blood pressure is-0.0042 mmHg, 0.3730 mmHg, and 0.2472 mmHg respectively, meeting the standards of Association for the Advancement of Medical Instrumentation and A-level British Hypertension Society. This study shows that the proposed sensor has broad application prospects in high-precision wearable medical electronics.