Abstract:In recent years, bile acids (BAs), the important component of bile, were found closely related to the oc-currence and development of diseases, therefore, determination of BAs in bile samples is of great signifi-cance. However, biological matrix complexity and low concentrations of BAs were still challenging for BA detection in small amount of bile samples. In this work, a core-shell NH2-MIL101@mSiO2 was designed to improve the capture ability of BAs in biological samples, as well as possess good biocompatibility. Sub-sequently, solid-phase microextraction (SPME) probe of the NH2-MIL101@mSiO2 was coupled with HPLC-MS/MS to establish the analysis method for detecting eight BAs in bile samples. The established method received extraction efficiencies of (30-2143)-fold higher than those of the commercial probes and low limit of detection (LOD <= 0.21 ng mL -1). The miniaturization of SPME sampling devices, as well as the low LOD of this work, endowed this method advantage of low consumption of bile samples (30 mu L). Based on the proposed method, eight BAs in bile samples of pancreatic cancer patients and cholelithiasis patients were detected successfully. A distinct difference was found in the concentrations of four tar-geted BAs in bile samples from pancreatic cancer patients and cholelithiasis patients. This work provided a method for quantification of eight BAs in small volume human bile samples, and it could open up a perspective regarding the relationship between BA metabolism and the occurrence of diseases.(c) 2022 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?