Single Wire Capacitive Wireless Power Transfer System for Wearable Biomedical Sensors Based on Flexible Graphene Film Material.

Lei Yang,Baoxiang Feng,Yuanqi Zhang,Xiaojie Li,Liqiang Zhang,Xinze Chen,Jiarui Nie,Haibing Wen,Jiaqiang Tian,Jiale Jian,Jingjing Huang,Darui Zhu,Aimin Zhang,Xiangqian Tong,Liaoyong Wen
DOI: https://doi.org/10.1109/tbcas.2022.3205762
2022-01-01
IEEE Transactions on Biomedical Circuits and Systems
Abstract:This paper provides a special flexible graphene film based capacitive wireless power transfer (FGCPT) system for powering biomedical sensors of smart wearable devices. The graphene conductive material is flexible, transparent, highly conductive, and impermeable to most gases and liquids. Generally, the coupling structure of capacitive wireless power transfer (CPT) system is consisted of metal plates. However, it is hard to use for the biomedical sensors as the low power density and big volume. The shape of graphene conductive material could be easily built and changed according to the application requirements. In this paper, the power supply of biomedical sensing system could be accomplished by a single graphene film which is acted as the receiver of FGCPT system. The 200 mW power level is achieved with the maximum 9 V output voltage. The theory and calculation are verified by the simulated and experimental results.
What problem does this paper attempt to address?