Comparison of Ring Pulse Oximetry Using Reflective Photoplethysmography and PSG in the Detection of OSA in Chinese Adults: A Pilot Study.

Rui Zhao,Jianbo Xue,Xueli Zhang,Maohuan Peng,Jing Li,Bing Zhou,Long Zhao,Thomas Penzel,Meir Kryger,Xiao Song Dong,Zhancheng Gao,Fang Han
DOI: https://doi.org/10.2147/nss.s367400
2022-01-01
Nature and Science of Sleep
Abstract:Objective A novel ring-worn oximeter (Circul) uses reflective photoplethysmography and automated signal processing to calculate oxygen desaturations. We evaluated the ability of Circul to detect obstructive sleep apnea in Chinese adults. Methods We recruited 207 Chinese Han subjects: 70% males, mean age 48.2±14.7 years, mean BMI 27.6±4.8 kg/m2 and mean AHI 28.6±25.2 events/h. All participants underwent simultaneous polysomnography (PSG) and Circul testing in a sleep laboratory. Oxygen desaturation index (ODI), mean oxygen saturation (MSpO2), cumulative time at SpO2<90% (CT90), cumulative percentage of sleep time spent with SpO2<90% (CT90/TST) were derived and compared for the Circul and the PSG. Results The ODI was 25.3±24.5 events/h using PSG and 22.2±24.5 events/h using Circul (P<0.0001), with an intraclass correlation coefficient (ICC) of 0.884. CT90 and CT90/TST between the two methods were not different; the MSpO2 level calculated by PSG was slightly lower than Circul, 95.0% (93.0–96.0%) vs 95.3% (93.9–96.6%), P<0.0001. Circul-ODI had a good correlation (r=0.91, p<0.0001) and close agreement with PSG-AHI (Bland-Altman analysis: Mean Difference 6.4, 95% CI −14.8 to 27.5 events/h). Using a threshold of AHI ≥5 events/h, the Circul had 87% sensitivity, 83% specificity, 5.09 positive likelihood ratio (LR+), 86% accuracy, and 0.929 area under the curve (AUC). Conclusion Circul ring pulse oximetry can detect OSA with reasonable reliability. The Circul system is a reliable and comfortable choice for OSA assessment.
What problem does this paper attempt to address?