A Novel 3-D Force Perception Structure Based on Fiber Bragg Grating for Surgical Robots

Leyang Wang,Yongzhuo Gao,Yuxin Zhang,Xiangrong Wang,Zhijiang Du
DOI: https://doi.org/10.1109/icma54519.2022.9856313
2022-01-01
Abstract:Due to the existing minimally invasive surgical robot’s ability to perceive external forces still needs to be improved, doctors can only estimate the size of the operating force through visual feedback when performing surgery through the robot, which makes the quality and safety of the operation greatly being affected. This article designed a novel 3D force perception system based on the Fiber Bragg Grating Sensor (FBG). It contains four fibers and one of them set at the center of sensor. The contribution of this novel structure has two holes on the central beam so it will be easier to fix the central fiber. Besides, a novel preloading device shorten the time of fixing apparently. This article analyzes the force perception model and a six DoFs experimental platform has been built for calibration. The performance of FBG sensor and temperature decouple algorithm has been verified. Finally, the sensitivities of the designed FBG sensor in the three-axis force direction are 87.50 pm/N, 76.20 pm/N, and 19.18 pm/N, respectively. Moreover the three-dimensional force can be measured in any spatial direction.