Low-cost high-resolution photoacoustic microscopy of blood oxygenation with two laser diodes

Xiufeng Li,Kylie Yeung,Victor T. C. Tsang,Bingxin Huang,Claudia T. K. Lo,Terence T. W. Wong
DOI: https://doi.org/10.1364/BOE.458645
IF: 3.4
2022-01-01
Biomedical Optics Express
Abstract:Optical-resolution photoacoustic microscopy (OR-PAM) has been widely used for imaging blood vessel and oxygen saturation of hemoglobin (So(2)), providing high-resolution functional images of living animals in vivo. However, most of them require one or multiple bulky and costly pulsed lasers, hindering their applicability in preclinical and clinical settings. In this paper, we demonstrate a reflection-mode low-cost high-resolution OR-PAM system by using two cost-effective and compact laser diodes (LDs), achieving microvasculature and so(2) imaging with a high lateral resolution of similar to 6 mu m. The cost of the excitation sources has dramatically reduced by similar to 20-40 times compared to that of the pulsed lasers used in state-of-the-art OR-PAM systems. A blood phantom study was performed to show a determination coefficient R-2 of 0.96 in linear regression analysis. Experimental results of in vivo mouse ear imaging show that the proposed dual-wavelength LD-based PAM system can provide high-resolution functional images at a low cost. (C) 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement
What problem does this paper attempt to address?