Laser-cooled <SUP>171</SUP>Yb<SUP>+</SUP> microwave frequency standard with a short-term frequency instability of 8.5 x 10<SUP>-13</SUP>/̦

S. N. Miao
DOI: https://doi.org/10.1364/OE.453423
IF: 3.8
2022-01-01
Optics Express
Abstract:We report on the development of a microwave frequency standard based on a laser-cooled Yb-171 (+ )ion trap system. The electronics , lasers, and magnetic shields are integrated into a single physical package. With over 10(5) ions are stably trapped, the system offers a high signal-to-noise ratio Ramsey line-shape. In comparison with previous work, the frequency instability of a Yb-171 (+) microwave clock was further improved to 8.5 x 10(-13) /root tau for averaging times between 10 and 1000 s. Essential systematic shifts and uncertainties are also estimated. (C) 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement
What problem does this paper attempt to address?