Design and finite element analysis of sandwich transducer for ultrasonic scalpel

LIU Yuyang,LIU Ping,CHEN Xiaohong,SUN Fujia,ZHOU Honglei
DOI: https://doi.org/10.13258/j.cnki.nmme.2022.03.003
2022-01-01
Abstract:The traditional transducer design method has low efficiency and high cost. A piezoelectric transducer for ultrasonic scalpel with 55 000 Hz was designed by using theoretical calculation and finite element method to improve the design efficiency and verify the feasibility of the method. The ultrasonic transducer was designed as a sandwich structure, and PZT-8 piezoelectric ceramics were used for design and simulation. Firstly, the initial size of each component of the transducer was theoretically designed and three-dimensional modeling was carried out. The model was imported into workbench software for modal analysis and harmonic response analysis, the resonant frequency and output amplitude of the transducer were extracted, and the final size of the transducer was determined. The solid transducer was processed and assembled according to the final size. Finally, the resonant frequency of the entity was measured by impedance analyzer. The results show that the resonant frequency of the simplified simulation model is approximately equal to the measured value, which is 54 980 Hz. It proves the rationality and feasibility of the design idea, and provides reference help for the working mode and optimal design of the energy converter.
What problem does this paper attempt to address?