Enhanced Sonothrombolysis Induced by High-Intensity Focused Acoustic Vortex

Shifang Guo,Zhen Ya,Pengying Wu,Lei Zhang,Mingxi Wan
DOI: https://doi.org/10.1016/j.ultrasmedbio.2022.05.021
IF: 3.694
2022-01-01
Ultrasound in Medicine & Biology
Abstract:High-intensity focused ultrasound (HIFU) thrombolysis provides a targeted and non-invasive therapy for thrombosis-related diseases. Rapid thrombolysis and restoration of blood flow are vital to reduce the disability and death rate. The objective of this study was to explore the feasibility of using a high-intensity focused acoustic vortex (HIFAV) to enhance sonothrombolysis. The in vitro clots were treated with HIFU with a peak negative pressure (PNP) of 2.86 MPa (HIFU A) or 3.27 MPa (HIFU B) or HIFAV with a PNP of 2.14 MPa. The results revealed that HIFAV thrombolysis could achieve a significantly higher efficiency than HIFU (HIFAV: 65.4%, HIFU A: 24.1%, HIFU B: 31.6%, p < 0.01), even at a lower intensity. The average size of the debris particles generated in HIFAV thrombolysis was similar to that in HIFU. Additionally, the cavitation activities were found to be more intense in HIFAV thrombolysis. Although the efficiency of HIFAV thrombolysis was higher when the pulse repetition frequency increased from 100 to 500 Hz (41.4% vs. 65.4%, p < 0.05), it decreased when the PRF reached 1000 Hz (29.9%). Lastly, it was found that increasing the duty cycle from 5% to 15% led to a higher efficiency in HIFAV thrombolysis (40.3% vs. 75.2%, p < 0.001). This study illustrated that HIFAV provided enhanced thrombolysis and that its efficiency could be further increased by optimizing the ultrasound parameters.