Research on Biomechanical Application of Fracture Reduction Fixator Based on Finite Element Analysis

Kai Wu,Jianxiong Ma,Yuren Du
DOI: https://doi.org/10.1088/1742-6596/1852/2/022088
2021-01-01
Journal of Physics Conference Series
Abstract:This article uses CAXA 3D solid design software to model the fracture reduction fixator, and assemble the model according to the requirements of clinical medicine. Obviously apply 50N, 100N, 150N, 200N axial load to the model, and obtain the relationship between axial load force and displacement. The stiffness of the model calculated by SPSS23.0 statistical analysis software is 88.5N/mm. The stiffness of the fracture reducer measured by the biomechanical model was 85.4N/mm, and the ratio between the two was 1.037, which was within the error range. Prove that the model and calculation method established in this study can provide reference for clinical medicine.
What problem does this paper attempt to address?