Modeling Calculation of Tumor Treatment Field in Human Thoracic and Optimization Strategy of Electrode Array

Haoxiang Lv,Shoulong Dong,Yancheng Wang,Wencheng Peng,Jianhao Lin,Liang Yu,Chenguo Yao
DOI: https://doi.org/10.1007/978-981-19-1532-1_130
2022-01-01
Abstract:The tumor treatment field imposes a low intensity (1–3 V/cm) intermediate frequency (100 kHz–500 kHz) alternating electric field on the tumor cells to destroy the normal mitotic process of tumor cells, thereby inhibiting the proliferation of tumor cells. Clinically, it has been shown that the tumor treatment field can effectively slow down the growth of tumors under the premise of non-invasiveness to patients, and only some patients have mild dermatitis. This paper designs a tumor treatment electrode array suitable for lung cancer, builds an actual human thoracic cavity simulation model, first explores the distribution of tumor treatment fields in different tissues of the thoracic cavity, and clarifies that the field strength of the tumor area can reach the treatment threshold. Subsequently, the electrode array layout is rotated, and the field strength increase of the tumor at different rotation angles is analyzed, and the rotating electrode array layout with the largest field strength increase is found. The research in this paper is the first to conduct a more in-depth array layout optimization for a human thoracic cavity model. The proposed electrode array layout optimization scheme has a reference value for the application of the tumor treatment field in lung cancer.KeywordsTumor treatment fieldTumor cellsSimulation calculationsAlternating electric field
What problem does this paper attempt to address?