Forward volume scattering function (0.03 -60 ) measured using an oblique-incidence particle sizer

Chaofan Wu,Bangyi Tao,Yaorui Pan,Haiqing Huang,Zhihua Mao,Delu Pan,Jilai Fan,JiMing Zhang,ShiZhe Chen
DOI: https://doi.org/10.1364/OE.454837
IF: 3.8
2022-01-01
Optics Express
Abstract:The forward volume scattering function (VSF) is an inherent optical property important in ocean lidar and underwater imaging and communication. The scattered power within 60 degrees contains >90% of total scattered power, making it essential for determining the asymmetry parameter g. Thus, the new oblique-incidence-design Bettersize BT-3000 particle sizer was utilized to measure forward VSF (0.03 degrees-60 degrees) synchronously. A double-exponential model was then used to construct the full-angle-range VSF (0 degrees-180 degrees). The g value calculated therefrom had an uncertainty of <1%. Calibration was implemented using 11-mu m beads alone, and the BT-3000's performance was validated. (C) 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement
What problem does this paper attempt to address?