Electrochemical Determination of Paracetamol Using MXene/single-walled Carbon Nanohorns Composite As Sensor

Wei Zhong,Jing Yang,Yansha Gao,Xiaoqiang Wang,Shuwu Liu,Mingfang Li,Limin Lu,Xuemin Duan
DOI: https://doi.org/10.20964/2021.11.39
IF: 1.541
2021-01-01
International Journal of Electrochemical Science
Abstract:In this study, single-walled carbon nanohorns (CNHs)-intercalated MXene (Ti3C2) (MXene/CNHs) composite was successfully synthesized, which was used as electrochemical sensor for paracetamol (PAR) detection. The composite of MXene/CNHs exhibits excellent conductivity, large specific surface area and superior electrocatalytic properties. Specifically, the embedding of the CNHs into MXene nanosheets successfully prevents the re-stacking of MXene and provides abundant active sites for catalytic PAR oxidation, which can availably improve the current signal of the electrochemical sensor. Under the optimum conditions, the designed MXene/CNHs electrode can be employed to detect PAR in a broad concentration range of 5.0 nM - 10.0 mu M with a low limit of detection (LOD) of 2.0 nM (S/N = 3). Additionally, the modified electrode exhibits high selectivity, good reproducibility and acceptable stability.
What problem does this paper attempt to address?